79
15 Technical data
15
Technical data
15.1
IO-Link interface
15.2
Analog inputs (sensor connection side)
Input for temperaure sensor
Input for CR conductivity sensor
Communication interface
IO-Link device V 1.1 (downward compatible to V 1.0)
Communication mode (data transfer rate)
COM 3 (230.4 kBaud)
IO Device Description (IODD)
The IODD can be localized via the "IODDfinder" on the
JUMO website in the product area for this device or at
www.io-link.com and downloaded.
Max. cable length acc. to IO-Link standard
20 m
Output mode
Switching output type
Transistor switching output can be configured as NPN, PNP
or Push/Pull
short-circuit proof
Yes (clocked)
resistant to overload
Yes
protected against polarity reversal
Yes
Ampacity of the switching outputs
100 mA in each case
Voltage drop of the switching outputs
max. 2 V in each case
Measuring range
Pt100
-50 to +250 °C
Pt1000
-50 to +250 °C
Connection types
2-wire/3-wire
Measuring accuracy
±0.25 % of MR
a
Ambient temperature influence
0.1 % / K
Period
500 ms
a
MR: measuring range span
Units
µS/cm
mS/cm
kΩ × cm
MΩ × cm
Display ranges
a
0.000 to 9.999
00.00 to 99.99
000.0 to 999.9
0000 to 9999
Temperature compensation
TC linear
b
for -50 to +100 °C and over 0.045 µS/cm
TC curve
for -20 to +150 °C and over 0.055 µS/cm
TDS
c
for -50 to +250 °C
Natural water DIN EN 27888 for 0 to 36 °C
Natural water with expanded range for 0 to 100 °C
ASTM D1125-95 neutral impurities for
0 to 100 °C and over 0.045 µS/cm
ASTM D1125-95 alkali and acid impurities for
0 to 100 °C and over 0.055 µS/cm
Содержание digiLine CR HT10
Страница 2: ......
Страница 6: ...Contents...
Страница 16: ...3 Device Description 16...
Страница 26: ...4 Identifying the device version 26...
Страница 50: ...10 Troubleshooting in the event of faults 50...
Страница 70: ...12 IO Link data 70...
Страница 78: ...14 Operation maintenance and care 78...
Страница 83: ......