9 Calibration
44
temperature coefficient for each interval. This requires installation of a temperature sensor that the de-
vice can use to sense the temperature of the sample solution. The series of temperature values consists
of 7 values in total:
•
"Starting temperature and end temperature of the TC-curve" (see "Calibrating the temperature co-
efficient curve (TC-curve)", page 47)
•
Reference temp. (see chapter 13.3.1 "CR input (resistivity conductivity)", page 71)
•
4 additional temperature values between the "starting temperature and end temperature of the TC-
curve"
The "end temperature of the TC-curve" be at least 20 °C higher than the "starting temperature of the TC-
curve". The two values are requested at the start of the TC-curve calibration and must be entered by the
user.
The reference temperature must lie between the "starting temperature and end temperature of the TC-
curve" and must differ from the starting temperature and end temperature by at least 1 °C. It is set in the
configuration of the Ci conductivity input.
The intervals between the starting, reference-and end temperature are automatically divided into 6 inter-
vals by the device. The remaining 4 temperatures are determined in this way.
C
o
n
duct
i
v
i
ty
6 cal
i
brat
i
o
n
po
in
ts
+
refere
n
ce temperature
6 I
n
tervals
T
emperature
Starting temperature
TC-Curve
Reference tempe
4 additional temperature values
End temperature
TC-Curve
Содержание digiLine CR HT10
Страница 2: ......
Страница 6: ...Contents...
Страница 16: ...3 Device Description 16...
Страница 26: ...4 Identifying the device version 26...
Страница 50: ...10 Troubleshooting in the event of faults 50...
Страница 70: ...12 IO Link data 70...
Страница 78: ...14 Operation maintenance and care 78...
Страница 83: ......