12 IO-Link data
56
Process data input 2 (process values as TINT32)
Designation
Data type Value range
Default
Description
Uncompensated
conductivity pro-
cess value
TINT32
Sensor measuring
range
0
This value must be divided by
1000 to obtain the process value.
Compensated
conductivity pro-
cess value
TINT32
Sensor measuring
range
0
Temperature pro-
cess value
TINT32
Sensor measuring
range
0
This value must be divided by 10
to obtain the process value.
Measuring range
selection
TUINT8
(bit field)
0x00 = MR 1
0x01 = MR 2
0x02 = MR 3
0x03 = MR 4
0
Status
TUINT16
(bit field)
Bit 0: Invalid conductivi-
ty process value
Bit 1: Invalid tempera-
ture process value
Bit 2: Compensation
temperature range left
Bit 3: Max. CIP/SIP cy-
cles pre-alarm
Bit 4: Max. CIP/SIP cy-
cles alarm
Bit 5: Sensor stress pre-
alarm
Bit 6: Sensor stress
alarm
Bit 7: Calibration timer
expired
Bit 8: Operating data ac-
quisition alarm
Bit 9: -
Bit 10 to Bit 13: reserved
Bit 14: Configuration
data error
Bit 15: Calibration data
error or device defective
0
Limit value switch
status
TUINT8
(bit field)
Bit 0: Limit value
switch 1
Bit 1: Limit value
switch 2
0
Содержание digiLine CR HT10
Страница 2: ......
Страница 6: ...Contents...
Страница 16: ...3 Device Description 16...
Страница 26: ...4 Identifying the device version 26...
Страница 50: ...10 Troubleshooting in the event of faults 50...
Страница 70: ...12 IO Link data 70...
Страница 78: ...14 Operation maintenance and care 78...
Страница 83: ......