11 Data overview
11.4
Operating data
CIP/SIP alarm signal
On reaching the maximum number of CIP or SIP cycles set for this alarm sig-
nal, this alarm is triggered. It can be retrieved by the IO-Link master.
The maximum number of CIP/SIP cycles for this alarm is set in the configu-
ration data.
chapter 13 "Configuration", page 71
Data point
Explanation
Data point
Explanation
Operating Hours Counter
The operating hours counter records the total operating time of the JUMO
digiLine electronics to the second. It cannot be configured or reset.
Sensor replacement counter
The sensor replacement counter records how often sensors on the JUMO
digiLineCRwere replaced. It provides a history on the PC of archived sensor
information and calibration logbook entries for every sensor with which the
JUMO digiLine electronics were operated. The sensor replacement counter
is incremented by the JUMO DSM software.
JUMO DSM software operating manual
Initial startup date
Initial startup date
Operating hours counter reading at
initial startup
Operating hours counter reading at startup on a JUMO digiLine master
Drag indicator for temperature
Lowest temperature value
Data on the highest or lowest temperature values that occurred inside the de-
vice in the course of operation to date.
Highest temperature value
Time of lowest temperature value
Time of highest temperature value
Operating hours counter reading at
lowest temperature
Operating hours counter reading
at lowest temperature
Extreme conditions
a
Total time below the minimum tem-
perature
Total time violating the admissible min./max. values of temperature and elec-
trolytic conductivity
The min./max. values of temperature and electrolytic conductivity depend on
the sensor used and are stored in the "Sensor information".
Total time above the maximum tem-
perature
Total time above the maximum con-
ductivity value
Number of times below the minimum
temperature
Number of times above or below the admissible min./max. values of tem-
perature and electrolytic conductivity
The min./max. values of temperature and electrolytic conductivity depend on
the sensor used and are stored in the "Sensor information".
Number of times above maximum
temperature
Number of times above the maximum
conductivity value
a
only available on device versions as head transmitter
Содержание digiLine CR HT10
Страница 2: ......
Страница 6: ...Contents...
Страница 16: ...3 Device Description 16...
Страница 26: ...4 Identifying the device version 26...
Страница 50: ...10 Troubleshooting in the event of faults 50...
Страница 70: ...12 IO Link data 70...
Страница 78: ...14 Operation maintenance and care 78...
Страница 83: ......