59
12 IO-Link data
Unit for calcula-
tion
128 to
131
for MR
1 to 4
1
TENUM
TENUM values:
0x00: µS/cm
0x01: mS/cm
0x02: kΩ × cm
0x03: MΩ × cm
0x01
RW
Compensation
2
TENUM
TENUM values:
0x00: None
0x01: TC linear
0x02: TC-curve
0x03: nat. waters
0x04: nat. waters ex-
tended
0x05: TDS
0x06: ASTM neutral
0x07: ASTM acidic
0x08: ASTM alkaline
0x01
RW
Temperature co-
efficient
3
TFLOAT
0 to 8 %/K
2.4
RW
Offset
4
TFLOAT
-9999 to +9999
0
RW
TDS factor
5
TFLOAT
0.01 to 2.00
0.67
RW
Unit
6
TSTRING
up to 5 text charac-
ters
"ppm"
RW
a
RW = Read and write access
RO = Read access only
WO = Write access only
CR input