6 Electrical connection
34
Sensor connection
A connecting cable with wire end ferrules on the sensor end is available for the JUMO digiLine CR in
device versions with a separate sensor (see accessories). This cable must be provided with a cable
socket suitable for the selected sensor and connected in accordance with the following table.
CAUTION!
The connecting cable must be routed directly to the transmitter, not over terminal blocks.
8-pole sensor connection
Pin
Wire color
Function
Symbol
1
Red
Outer electrode 1
Socket
Connection side
2
Gray
Inner electrode 1
3
Yellow
Temperature 1
4
Green
Temperature 2
5
Pink
Inner electrode 2
6
blue
Outer electrode 2
7
Brown
Temperature 3
8
White
Shield
2
1
3
4
5
6
7
8
1
2
3
4
5
6
7
8
5-pole sensor connections
Sensor connecting cable 8-pole
Sensor-side connections
Pin
Wire color
Function
Angular connec-
tor on the sen-
sor
Fixed cable on
the sensor
M12 connector
5-pole on the
sensor
1
Red
Outer electrode 1
White
1
2
Gray
Inner electrode 1
3
Yellow
Temperature 1
Yellow
a
3
b
4
Green
Temperature 2
Green
5
Pink
Inner electrode 2
2
Brown
2
6
blue
Outer electrode 2
7
Brown
Temperature 3
-
8
White
Shield
-
Shield
-
a
2-wire circuit
b
3-wire circuit
Содержание digiLine CR HT10