13 Configuration
74
Configuration item
Selection/setting op-
tion
Explanation
CIP/SIP alarm
Inactive
Active
Activates/deactivates the CIP/SIP alarm
on reaching the maximum number of CIP/SIP cycles
Sensor stress alarm
Inactive
Active
Activates/deactivates the sensor stress alarm
The "Sensor stress" value reflects the current level of stress on
the sensor from high temperatures and high conductivity val-
ues. The following sensor stress alarm states are signaled on
JUMO masters and on the device display (only on device ver-
sions with a display) on reaching the specified limit values:
•
Pre-alarm for sensor stress above sensor stress level 3
•
Alarm for sensor stress above sensor stress level 7
CIP temperature
-20 to+150 °C
Temperature thresholds for identifying CIP/SIP cycles
If the CIP/SIP cycle takes place above one of these values for
the set duration of a CIP/SIP cycle, the values identify a suc-
cessfully completed CIP/SIP cycle and the CIP or SIP counter
is incremented. The respective counter is reset only after the
value has dropped below the CIP/SIP temperature.
SIP temperature
CIP cycle duration
0 to 9999 s
Duration of a CIP/SIP cycle
SIP cycle duration
Maximum number of
CIP cycles
0 to 999
Specifies the number of CIP/SIP cycles at which the CIP/SIP
alarm is triggered on the master
a
Maximum number of
SIP cycles
a
The counters for CIP and SIP cycles are automatically incremented by the JUMO digiLine electronics each time a
CIP or SIP process is recognized on the basis of the configured CIP/SIP temperatures and duration of the CIP/SIP
cycle.
Содержание digiLine CR HT10
Страница 2: ......
Страница 6: ...Contents...
Страница 16: ...3 Device Description 16...
Страница 26: ...4 Identifying the device version 26...
Страница 50: ...10 Troubleshooting in the event of faults 50...
Страница 70: ...12 IO Link data 70...
Страница 78: ...14 Operation maintenance and care 78...
Страница 83: ......