13 Eliminating faults and malfunctions
64
13.1.2 Reference liquid test
Immerse in test
solution
Test
sequence
✱
Prepare the conductivity test solution in a container of adequate size.
✱
Wire up the device, see Chapter 7 "Installation", Page 25.
✱
Select the range appropriate to the conductivity test solution, see Chapter
10.6.1 "CONDUCTIVITY IN (Conductivity input)", Page 37 -> RANGE 1 — 4
✱
Set TC to 0%/°C, see Chapter 10.6.1 "CONDUCTIVITY IN (Conductivity
input)", Page 37 -> TEMPCO
✱
Immerse the cell in the container, and do not move it any more during the
measurement.