53
11 Calibration
- The sensor must be connected to the transmitter (applies to the split
version).
- The transmitter is in the measurement mode.
✱
Immerse the conductivity sensor in a sample of the solution to be
measured.
✱
Press the
key for at least 3 seconds.
✱
Use the
and
keys to select CALIBRATION LEVEL; use the
key to
confirm the selection.
✱
Use the
and
keys to select TEMPCO LINEAR; use the
key to
confirm the selection.
✱
Use the
and
keys to enter the working temperature; confirm with the
key.
The working temperature must be at least 5°C above or below the
reference temperature (25.0°C).
PGM
PGM
PGM
PGM