13 Eliminating faults and malfunctions
Error
possibilities
t
13.1 Checking the device
General
The device is calibrated at the factory, and is maintenance-free. If,
nevertheless, measurement deviations appear with no apparent cause, the
transmitter can be tested as follows.
Problem
Possible cause
Measures
no measurement display
or
current output
supply voltage
missing
supply voltage
should be checked,
also check terminals
measurement display
000 or
current output 4 mA
sensor not immersed in
medium,
reservoir level
too low
top up the reservoir
flow-through fitting
is blocked
flow-through fitting
should be cleaned
sensor is faulty
see "Checking the
device", Page 61
wrong or
unstable
measurement display
sensor not immersed
deeply enough
top up the reservoir
inadequate mixing
ensure good
mixing
for sensor: all-round
free space of approx.
5 mm
ensure all-round flow
air bubbles
check mounting site,
Measurement value dis-
play 8888,
temperature display
"ok",
blinking
Conductivity measure-
ment range overshoo-
ting or conductivity mea-
surement probe faulty.
Select suitable measure-
ment range.
Replace conductivity
measuring transmitter.
Measurement value dis-
play 8888,
temperature display
8888
blinking
Temperature measure-
ment range overshoo-
ting or undershooting or
short circuit or interrupti-
on of the temperature
sensor.
The temperature of the
measuring medium must
be in the range from
0...150°C.
Replace the conductivity
measuring transmitter.
Send the device for re-
pairs.
Содержание 202755/20
Страница 19: ...19 6 Mounting 6 2 Head mounted transmitter Installation variations...
Страница 20: ...6 Mounting 20 6 3 Split version separate sensor Transmitter head Drilling jig for wall mounting...
Страница 21: ...21 6 Mounting Sensor component...
Страница 35: ...35 10 Operation 10 4 Principle of operation Operation in levels...
Страница 71: ......