29
8 Setup program
8.1
Function
Configurable
parameters
The setup program, which is available as an option, can be used for easy
adaptation of the transmitter to specific requirements.
- Setting the measurement range and the range limits.
- Setting the response of the output to an out-of-range signal.
- Setting the functions of the switched outputs K1 and K2.
- Setting the functions of the binary inputs E1 and E2.
- Setting up special functions (e.g. the dilution function).
- Setting up a customer-specific characteristic,
- etc.
Connections
Data transmission from or to the transmitter can only take place
when it is connected to the electrical supply see Chapter 7
"Installation", Page 25ff.
The setup interface is not electrically isolated. When connecting
the PC interface cable, it is therefore absolutely essential to ensure
that either the supply of the transmitter or of the PC is not
electrically earthed (for instance, use a battery-powered notebook).
Interface cable
(95/00350260)
Plug-on adapter
(included in setup set)
Supply voltage