10 Operation
38
CUSTOMIZED
(values can only be entered by using the optional
setup program)
CONC. MEAS.
RANGE
For HNO
3
0 — 25 % BY WEIGHT
36 — 82 % BY WEIGHT
For NaOH
0 — 15 % BY WEIGHT
25 — 50 % BY WEIGHT
OFFSET
-100 to 0 to +100 mS/cm (+/- 10% of range)
FILTER TIME
00:00:00 — 00:00:01 — 00:00:25 H:M:S
CALIB. INTERVAL
0 — 999 DAYS (0 = switched off)
10.6.2 CONDUCTIVITY OUT (conductivity output)
SIGNAL TYPE
0 — 20 mA
4 — 20 mA
20 — 0 mA
20 — 4 mA
0 — 10 V
2 — 10 V
10 — 0 V
10 — 2 V
SCALING START 1 — 4
1
0 µS/cm = 4 mA
Can be set in the range being used, depending on the signal type.
1
Ranges 2, 3 and 4 are only used if
BINARY INPUT is configured to RANGE/TEMPCO.
SCALING END 1 — 4
1
1000 µS/cm = 20 mA
Can be set in the range being used, depending on the signal type.
1
Ranges 2, 3 and 4 are only used if
BINARY INPUT is configured to RANGE/TEMPCO.