10 Operation
10.6.1 CONDUCTIVITY IN (Conductivity input)
RANGE 1 — 4
1
0 — 500 µS/cm
0 — 1000 µS/cm
0 — 2000 µS/cm
0 — 5000 µS/cm
0 — 10 mS/cm
0 — 20 mS/cm
0 — 50 mS/cm
0 — 100 mS/cm
0 — 200 mS/cm
0 — 500 mS/cm
0 — 1000 mS/cm
0 — 2000 mS/cm UNC
2
1
Measurement ranges 2, 3 and 4 are only used if
BINARY INPUT is configured to RANGE/TEMPCO
2
This measurement range is not temperature-compensated.
TEMP. COMPENSATION
LINEAR
NON-LINEAR (see "Non-linear temperature coefficient (ALPHA)", Page 55)
NATURAL WATER (permissible temperature range 0 to 36°C to EN 27 888)
TEMPCO 1 — 4
1
0 — 2.20 — 5.5%
1
Ranges 2, 3 and 4 are only used if
BINARY INPUT is configured to RANGE/TEMPCO.
REFERENCE TEMP.
15.0 to 25.0 to 30°C
REL. CELL CONSTANT
80.0 — 100.0 — 120%
MOUNTING FACTOR
80.0 — 100.0 — 120%
If it is not possible to achieve the minimum clearance of 20 mm between the
sensor and the outer wall, then a limited compensation can be made through
this parameter.
CONC. MEAS. TYPE
NO FUNCTION
NaOH
HNO3