57
11 Calibration
Requirements
- The supply voltage for the device must be present.
see Chapter 7 "Installation", Page 25ff.
- The sensor must be connected to the transmitter (applies to the split
version).
- The transmitter is in the measurement mode.
✱
Immerse the conductivity sensor in a sample of the solution to be
measured.
✱
Press the
key for at least 3 seconds.
✱
Use the
and
keys to select CALIBRATION LEVEL; use the
key to
confirm the selection.
✱
Use the
or
key to select TEMPCO NON-LIN.; use the
key to
confirm the selection.
The rate of change of the temperature must not exceed
- 10°C / min. for an exposed temperature sensor, and
-
0
1°C / min for an internal temperature sensor.
PGM
PGM
PGM
Содержание 202755/20
Страница 19: ...19 6 Mounting 6 2 Head mounted transmitter Installation variations...
Страница 20: ...6 Mounting 20 6 3 Split version separate sensor Transmitter head Drilling jig for wall mounting...
Страница 21: ...21 6 Mounting Sensor component...
Страница 35: ...35 10 Operation 10 4 Principle of operation Operation in levels...
Страница 71: ......