System Specifications
ITEM
SPECIFICATION
COMMENT
Environment
Indoor use at altitudes
up to 2000 m, Pollution
Degree 2
20% to 80% RH.
-
Operating Temperature
5 °C to 25 °C
-
RMS Noise Performance
Typical < 10 pA
For +/- 200 nA range
NP200
Data Sampling Rate
50,000 samples per second -
USB
Standard
connector
USB Mini-B
Cable must be to a relevant
safety standard such as IEC
60950-1 for IT equipment
or IEC 61010-1 for
laboratory equipment and
should provide double or
reinforced insulation from
hazardous voltage sources
Max input
voltage
5.00 ± 0.25 V
Electrical
Power
Voltage
12 V (DC)
At input jack on instrument
Current
130 mA
Power
consumption
1.5 W
PSU
Input AC
90 to 264 Vac 50/60
Hz ac
Use only Izon supplied
TRG45A120-21E11
Output DC 12 V nom
Output
current max
3.75 A
Output
power max
45 W
2
Nanopore holding stage
extension range
Min. 41 mm
Max. 61 mm
Measured across the outside
of opposing holding arms
Electrical safety EMC
IEC/EN 61010-1
IEC/EN 61326-1
-
1 Under electrical interference noise may increase but will return to normal operation on end of
interference. Position equipment away from electrical switching gear and interfering equipment or noise
may increase under sample running conditions.
2 PSU module rating. Izon Instruments running normally are approx. less than 1.6 W.
9
www.izon.com
Содержание qNano
Страница 1: ...www izon com qNano USER MANUAL...
Страница 4: ...GETTING STARTED 4 www izon com...
Страница 10: ...THEORY OF OPERATION 10 www izon com...
Страница 25: ...MEASUREMENT PLANNING 25 www izon com...
Страница 28: ...SAMPLE PREPARATION 28 www izon com...
Страница 34: ...INSTRUMENT OPERATION 34 www izon com...
Страница 47: ...TROUBLESHOOTING 47 www izon com...
Страница 56: ...www izon com...