Resolution range
When using the custom planner, the user will be asked to enter measurement
details, one of which will be the resolution range:
The resolution range is an indication of the size range that can be observed at
a single stretch. The default resolution range for each nanopore is calculated
at mid-range stretch (e.g. the NP250 default range at 45-46 mm is ~135-
540 nm).
The smallest particle can be adjusted within the nanopore size range to suit
your sample. Once this has been edited, the largest detectable particle size
will be calculated automatically, but cannot be larger than the upper range of
the nanopore (e.g. 630 nm for NP250 at mid-range stretches).
The resolution range ensures that the software guides the user to the
optimised instrument setting for the selected range of particles. It also allows
the user to measure the same size range every time regardless of instrument
use or nanopore size.
21
www.izon.com