6 / Optimise signal-to-noise ratio for calibration particles
The standard sample changeover protocol should be used any time particles
or samples are being added or changed.
—
Remove upper fluid cell and immediately add a droplet of electrolyte
onto the nanopore.
—
Rinse the inside of the upper fluid cell with deionised water and dry with
compressed air.
—
Gently wipe the top of the nanopore dry with a lint free tissue, fit the
upper fluid cell and load 35
µ
L of the next sample before connecting the
PM2 nozzle and applying your chosen pressure.
In order to obtain high quality data, certain parameters can be optimised.
These are the stretch, the Voltage, and the Pressure.