If working with biological samples, ensure that the samples are processed
thoughtfully, to preserve sample integrity and optimise the yield of target
particles.
Pore set-up, signal optimisation, and calibration
Mastering the reverse pipetting technique is essential when working with the
instrument. Using normal pipetting techniques can introduce bubbles into the
system, which can prevent the baseline current from being established.
Another common issue that can prevent the baseline current from being
established is users re-fitting the fluid cell when it has not been completely
dried. Whenever the fluid cell is removed it must be firmly patted dry. When
washing the fluid cell, only wash the inner chamber, do not immerse or wash
the outside. The most common cause of a wandering baseline is a fluid cell
that has fluid trapped inside it.
The easiest way to restore the function of a nanopore is to thoroughly wash
it with deionised water (using the procedure shown earlier). Flush the pore as
shown, then re-coat and re-equilibrate the pore.
If you are working with exosomes or very small particles (> 150 nm), you should
filter samples through a 0.22 µm filter.
REPAIR AND SERVICING OF YOUR IZON INSTRUMENT
There are no user-serviceable parts. Return equipment to Izon for service.
All replacement parts, cables, mains lead, power supply, Nanopores, Fluid
Cells must be obtained via Izon Science Ltd.
5 /
53
www.izon.com
Содержание qNano
Страница 1: ...www izon com qNano USER MANUAL...
Страница 4: ...GETTING STARTED 4 www izon com...
Страница 10: ...THEORY OF OPERATION 10 www izon com...
Страница 25: ...MEASUREMENT PLANNING 25 www izon com...
Страница 28: ...SAMPLE PREPARATION 28 www izon com...
Страница 34: ...INSTRUMENT OPERATION 34 www izon com...
Страница 47: ...TROUBLESHOOTING 47 www izon com...
Страница 56: ...www izon com...