Calibrating particle concentration
Particle concentration is proportional to the change in blockade rate per unit
of applied pressure. Accurate concentration values are typically derived at 2
or more pressure steps using calibration particles of a known concentration.
Figure 6. The particle rate observed should increase linearly with applied pressure:
This rate plot is used to check for system stability. This is crucial as the
concentration value is measured by comparing the gradient of the rate plot of
sample to calibration measurements.