ESSENTIAL CONSIDERATIONS
1.
Ensuring the suspension conditions are suitable for TRPS analysis
2.
Excluding contaminants from the suspension
3.
Maintaining the structural and functional integrity of target sample
particles
4.
Enhancing suspension stability
Use the TRPS measurement checklist to ensure you have the equipment
you need.
1 / TRPS suspension requirements
General points
As TRPS is a conductance-based technique, all sample solutions must have
an ionic strength of at least 0.01 M. Samples dissolved in pure water can be
spiked with small amounts of concentrated PBS to rectify this issue. However,
the requirement for solutions to be conductive generally excludes samples
that are solvent based.
Contamination
Besides preventing sample degradation, mitigating contamination of samples
is required to maintain system stability during recordings. Small nanopores
especially are easily blocked by large precipitates, aggregates, and microbial
contaminants. All samples and reagents that come into contact with
nanopores must be filtered with a 0.22 µm syringe filter daily.