RECOVERING AN UNSTABLE BASELINE AND
MAINTAINING PORE STABILITY
—
Gently tap the fluid cell cap to dislodge bubbles on top of the pore, and
mix the sample using the reverse pipetting technique, without introducing
more bubbles.
—
Replace the fluid in the lower fluid cell periodically (once every half-hour)
to prevent bubbles from gathering around the pore.
—
If the baseline is very unstable, there may be many invisible bubbles
beneath the pore. Put a droplet of electrolyte on top of the pore, take
the nanopore off the qNano and wipe the bottom gently with a lint free
tissue before remounting on the qNano the applying the same stretch as
before. Re-establish your baseline current.