TDX WHEELCHAIRS
32
Part No. 1114809
SECTION 3
WHEELCHAIR OPERATION
WHEELCHAIR OPERATION
Emergency Stop (Reset) Connector -
The connector accepts a 1/8-inch diameter
Phono plug. The Emergency Stop switch must be an open contact for normal driving
and a closed contact to activate the Emergency Stop function.
PIN
DESIGNATION
TIP
RESET (EMERGENCY STOP)
RING
COMMON (B)
REMOTE ON/OFF
- The remote on/off switch input allows the power switch to be
operated by an ability switch (normally open momentary switch with mono plug). To
use the remote on/off feature, the Drive Select/On/Off switch must be in the ON
position. Each activation of the ability switch will alternately turn the joystick ON or
OFF.
If a fault is detected, the cause of the fault will be scrolled across the second line of the
display.
Emergency Stop Switch (Reset) -
The emergency stop switch is used to stop the
chair and to select the operating mode for the chair. The switch input is located on the
control module next to the joystick input connector. An emergency stop switch is
needed whenever any of the following operating modes are programmed.
Environmental Controls (E.C.U.)
including recliner controls
3 Speed Mode in Momentary
Latched Modes
Pneumatic Control
Stand-by Mode
RIM Control
Remote Drive Selection Mode
Information Center Display Selection (does not require Reset activation at
power up)
If any of the above modes are selected, the control will require activation of the switch
immediately after the power switch is turned on in order to enter the drive mode. The
second line of the LCD will display -PRESS RESET.