Invacare® Platinum® Mobile
General
Alarm
Name
Alarm
Code
Alarm Code Name
Display Message
Possible Reason
Solution
Depleted
Battery
0x1100
SDA_depletedBattery
The display shows:
"Shut Down Battery
Depleted"
1.
The concentrator is
running off battery
power and the
battery charge
level is less than
or equal to 5%.
1.
Supply unit with
external power
(AC/DC) or install
a charged battery
into the unit.
2.
If alarm persists:
a. Inspect
the battery
connector for
debris.
b. Reinstall
battery.
3.
If alarm persists,
contact Invacare
for service.
Battery Too
Hot (shut
down)
0x1101
SDA_batteryTooHot
The display shows:
"Shut Down Battery
Too Hot"
1.
The concentrator is
running off battery
power and at least
one of the battery
temperatures is
greater than or
equal to 60°C.
1.
Remove the hot
battery and allow
to cool.
2.
Change the
battery.
3.
If alarm persists,
contact Invacare
for service.
30
1187375-B
All manuals and user guides at all-guides.com