Invacare® Platinum® Mobile
General
Alarm
Name
Alarm
Code
Alarm Code Name
Display Message
Possible Reason
Solution
No Breath
Detected
(Running)
0x3201
LPA_noBreathDetected
The display shows:
"No Breath Detected
Check Cannula" (not
displayed if in "Warm
Up" or "Prolonged
Warm Up")
1.
The concentrator
has not detected
an inhalation for
more than 15
seconds. -or-
2.
Restriction after
the conserver
valve.
1.
Verify cannula is
connected, not
kinked, properly
positioned and
user is breathing
through his/her
nose.
2.
Check the outlet
filter/Replace filter.
3.
If alarm persists,
contact Invacare
for service.
Breath Rate
Too High
(Running)
0x4202
LPA_breathRateTooHigh
The display shows:
"Breath Rate High
Reduce Activity"
1.
The user breath
rate has exceeded
the capacity of
the concentrator
for more than 15
seconds.
1.
Make sure breath
rate is lower than
40BPM.
2.
If alarm persists,
contact Invacare
for service.
24
1187375-B