Troubleshooting
General
Alarm
Name
Alarm
Code
Alarm Code Name
Display Message
Possible Reason
Solution
Low Battery
(~30
minutes
left)
0x7207
LPA_lowBattery30MinLeft
The display shows:
"Low Battery Charge
Unit"
1.
The concentrator is
running off battery
power and the
battery charge
level is calculated
to only be able to
provide between
15 to 30 minutes
of additional
operation before
reaching 0% state
of charge.
1.
Supply unit with
external power
(AC/DC) or install
a charged battery
into the unit.
2.
If alarm persists,
contact Invacare
for service.
Low Battery
(~15
minutes
left)
0x2200
LPA_lowBattery15MinLeft
The display shows:
"Low Battery Charge
Now"
1.
The concentrator
is running off
battery power
and the battery
charge level is
calculated to only
be able to provide
15 minutes or
less of additional
operation before
reaching 0% state
of charge.
1.
Supply unit with
external power
(AC/DC) or install
a charged battery
into the unit.
2.
If alarm persists,
contact Invacare
for service.
1187375-B
25
All manuals and user guides at all-guides.com