Troubleshooting
General
Alarm
Name
Alarm
Code
Alarm Code Name
Display Message
Possible Reason
Solution
Battery
Cold
(Running)
0x6206
LPA_batteryCold
The display shows:
"Battery Cold Allow to
Warm"
1.
The concentrator is
running off battery
power and at least
one of the battery
temperatures is
less than or equal
to –15° C
1.
Remove the cold
battery and allow
to warm.
2.
If alarm persists,
contact Invacare
for service.
Unit Hot
(Running)
0x5203
LPA_unitHot
The display shows:
"Unit Hot Allow to
Cool"
1.
The concentrator is
running off battery
power or external
power (AC/DC) and
the unit internal
temperature is
greater than or
equal to 50° C.
1.
Allow the unit to
cool.
2.
Clean air intake
filter.
3.
If alarm persists,
contact Invacare
for service.
Unit Cold
(Running)
0x5204
LPA_unitCold
The display shows:
"Unit Cold Allow to
Warm"
1.
The concentrator is
running off battery
power or external
power (AC/DC) and
the unit internal
temperature is less
than or equal to
5°C.
1.
Allow unit to
warm.
2.
If alarm persists,
contact Invacare
for service.