Invacare® Platinum® Mobile
General
Alarm
Name
Alarm
Code
Alarm Code Name
Display Message
Possible Reason
Solution
Battery
Too Hot
To Charge
(Running)
0xb20
b
LPA_batteryTooHot-
ToCharge
The display shows:
"Cannot Charge Allow
to Cool"
1.
The concentrator
is running off
external power
(AC/DC) and
at least one
of the battery
temperatures is
greater than or
equal to 49° C.
1.
Remove the hot
battery and allow
to cool.
2.
If alarm persists,
contact Invacare
for service.
Battery
Too Cold
To Charge
(Running)
0xb20c
LPA_batteryTooCold-
ToCharge
The display shows:
"Cannot Charge Allow
to Warm
1.
The concentrator
is running off
external power
(AC/DC) and
at least one
of the battery
temperatures is
less than or equal
to 8° C.
1.
Remove the cold
battery and allow
to warm.
2.
If alarm persists,
contact Invacare
for service.
Battery Hot
(Running)
0x6205
LPA_batteryHot
The display shows:
"Battery Hot Allow to
Cool"
1.
The concentrator is
running off battery
power and at least
one of the battery
temperatures is
greater than or
equal to 55°C.
1.
Remove the hot
battery and allow
to cool.
2.
If alarm persists,
contact Invacare
for service.
26