Troubleshooting
General
Alarm
Name
Alarm
Code
Alarm Code Name
Display Message
Possible Reason
Solution
Internal
Fault
0x1116
fanIsNotMovingAlarm
The display shows:
"Shut Down Internal
Fault"
1.
Cooling fan or PCB
fault.
1.
If alarm persists,
contact Invacare
for service.
Internal
Fault
0x1117
fanDriverAttemptI2cBus-
TimeOutAlarm
The display shows:
"Shut Down Internal
Fault"
1.
PCB or software
fault.
1.
If alarm persists,
contact Invacare
for service.
Internal
Fault
0x1119
serialeeTimeOutAlarm
The display shows:
"Shut Down Internal
Fault"
1.
PCB or software
fault.
1.
If alarm persists,
contact Invacare
for service.
Internal
Fault
0x111a
serialeeWriteNotVeri-
fiedAlarm
The display shows:
"Shut Down Internal
Fault"
1.
PCB or software
fault.
1.
If alarm persists,
contact Invacare
for service.
Internal
Fault
0x111b
serialeeErrorWriteAlarm
The display shows:
"Shut Down Internal
Fault"
1.
PCB or software
fault.
1.
If alarm persists,
contact Invacare
for service.
Internal
Fault
0x111c
errFifoEmptyOnGet
The display shows:
"Shut Down Internal
Fault"
1.
PCB or software
fault.
1.
If alarm persists,
contact Invacare
for service.
Internal
Fault
0x111d
lcdErrAllocatingMessage
The display shows:
"Shut Down Internal
Fault"
1.
PCB or software
fault.
1.
If alarm persists,
contact Invacare
for service.
Internal
Fault
0x111e
lcdErrSendingMessage
The display shows:
"Shut Down Internal
Fault"
1.
PCB or software
fault.
1.
If alarm persists,
contact Invacare
for service.