Troubleshooting
General
Alarm
Name
Alarm
Code
Alarm Code Name
Display Message
Possible Reason
Solution
Stuck
Button
0x1107
SDA_stuckButton
The display shows:
"Shut Down Stuck
Button"
1.
One of the
buttons on the
concentrator's
control panel was
pressed when a
power source is
being connected
to the unit.
1.
Make sure no
control panel
buttons are
pressed.
2.
Disconnect all
power sources
then reconnect to
a power source.
3.
If alarm persists,
contact Invacare
for service.
No Breath
Detected
(shut down)
0x110c
SDA_noBreathDetected
The display shows:
"Shut Down No Breath
Detected"
1.
An inhalation has
not been detected
by the unit for
more than 120
seconds.
1.
Verfy cannula is
connected, not
kinked, properly
positioned and
user is breathing
through his/her
nose.
2.
Check the outlet
filter/Replace filter.
3.
If alarm persists,
contact Invacare
for service.
Internal
Fault
0x1109
ricEventQueueFullAlarm
The display shows:
"Shut Down Internal
Fault"
1.
PCB or software
fault.
1.
If alarm persists,
contact Invacare
for service.