Invacare® Platinum® Mobile
General
Alarm
Name
Alarm
Code
Alarm Code Name
Display Message
Possible Reason
Solution
Internal
Fault
0x1120
highPressureAlarm
The display shows:
"Shut Down Internal
Fault"
1.
Blockage in outlet
gas stream.
Blockage may
be internal or
external.
1.
Remove any kinks
or blockage in
cannula.
2.
Replace the
patient outlet filter.
3.
If alarm persists,
contact Invacare
for service.
Internal
Fault
0x1121
lowPressureAlarm
The display shows:
"Shut Down Internal
Fault"
1.
Internal leakage of
pneumatic circuit.
2.
Blockage of inlet
flow.
1.
Clean air intake
filter screen.
2.
If alarm persists,
contact Invacare
for service.
Internal
Fault
0x1122
compressorRunningAlarm
The display shows:
"Shut Down Internal
Fault"
1.
Compressor or
compressor's
wire harness is
damaged. -or-
2.
The compressor
wire harness is
damaged or loose.
1.
If alarm persists,
contact Invacare
for service.
Internal
Fault
0x1123
compressorNotRun-
ningAlarm
The display shows:
"Shut Down Internal
Fault"
1.
Compressor or
compressor's
wire harness is
damaged. -or-
2.
The compressor
wire harness is
damaged or loose.
1.
If alarm persists,
contact Invacare
for service.