Part No. 715-0088, Rev. B
- 89 -
2.
Foreign material may have been drawn into instrument and
has blocked the sampling system. Check and clean if
necessary.
3.
Pressure transducer that detects plugged lines may have
failed. Replace power board.
14.
Mass flow display does not return to zero when no flow is present.
Possible solutions:
1.
There is no adjustment for the mass flow sensor. Flow 0.01
L/min is acceptable, however this may indicate an issue if it
does not return to zero. Run the MFM Flow Check/Leak
Check Procedure and verify MFM accuracy from 0-2.0 L/min.
Replace as required.
15.
SAMPLE BLOCKED not displayed when sample line intentionally blocked.
Possible solutions:
1.
Tube to pressure sensor disconnected or plugged into the
wrong port. Note this will also drastically lower the NO, NO
2
,
and O
2
sensors responses.
All routine maintenance, repairs, and replacement of standard parts should be conducted according to
procedures outlined in the Service Manual. For all clinical or technical issues not addressed in the
manual, please contact International Biomedical directly at:
Call 1-512-873-0033
FAX: 1- 512-873-9090
6.3. Customer Service
Should you need clinical or technical information, please contact International
Biomedical. To help us to help you, please have the following information available:
A complete description of the problem / observation
Model #
Serial #
Your Institution address
Your fax #
The above information is necessary to determine warranty status as well as to gather the
information necessary for us to fill out a Return Merchandise Authorization (RMA) should
you need to send anything back for repair.
Any equipment sent to International
Biomedical without an RMA will not be accepted for delivery.
Do not ship any
products to International Biomedical without first obtaining an RMA.
Содержание AeroNO 2.0
Страница 1: ......
Страница 3: ...This page intentionally left blank ...
Страница 6: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 3 SCLAIM MER ...
Страница 7: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 4 SCLAIM MER ...
Страница 8: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 5 SCLAIM MER ...
Страница 9: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 6 SCLAIM MER ...
Страница 10: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 7 SCLAIM MER ...
Страница 11: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 8 SCLAIM MER ...
Страница 12: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 9 SCLAIM MER ...
Страница 26: ...Par rt No 715 008 2 13 2 14 88 Rev B Physical D VESA Mou Dimensions unting Patte and Cente ern 23 er of Mass ...
Страница 78: ...Par rt No 715 008 88 Rev B 5 12 3 Re 5 12 4 Fr ear Interna ront Intern al Layout al Layout 75 ...
Страница 94: ......
Страница 95: ...Par rt No 715 008 7 3 88 Rev B Nitric Oxid de Sensor S Specificatio 92 ns ...
Страница 96: ...Par rt No 715 008 88 Rev B 93 ...
Страница 97: ...Par rt No 715 008 7 4 88 Rev B NO2 Senso or Specifica ations 94 ...
Страница 98: ...Par rt No 715 008 88 Rev B 95 ...
Страница 99: ...Par rt No 715 008 7 5 88 Rev B Oxygen Se ensor Spec cifications 96 ...
Страница 105: ......