Part No. 715-0088, Rev. B
- 13 -
Even when the external power supply is not connected, the battery provides a bias charge to the NO,
NO
2
, and O
2
cells to keep sensors in a ready state. The power drawn from biasing the sensors will
deplete a new fully charged battery in approximately one week. Because of this, it is recommended
the external power supply of the AeroNOx 2.0 remain plugged in at all times to maintain battery
charge.
Disconnect the battery or leave charging when not in use for extended periods of time.
Allowing the battery to fully discharge repeatedly will reduce the overall number of cycles in the life of
the battery.
Attempting to run the AeroNOx 2.0 for longer than five hours on battery only could result in an
interruption of Nitric Oxide therapy.
The specified power supply is defined as part of the ME Equipment. Use only the International
Biomedical approved universal power supply for AeroNOx 2.0 .
No other devices are known to cause potential interference with the AeroNOx 2.0 . If interference
occurs, discontinue use of the AeroNOx 2.0 and use the INOstat Bagger.
Never turn on the NO delivery gas without first turning on the ventilator or bagger flow. Failure to do
this will result in undiluted gas delivery entering the sampling chamber and exposing the sensors to
levels of NO and /or NO
2
which may damage the sensors.
NO
2
gas may have collected in the AeroNOx 2.0 delivery or ventilator circuit during setup. Run
ventilator and AeroNOx 2.0 on a test lung 30 seconds prior to patient connection to ensure analyzed
levels of NO
2
and NO are appropriate.
Only use the NO Worries Sample line with filter and Nafion tubing when operating the AeroNOx 2.0 .
Failure to do so may result in the egress of moisture which can impair function as well as damage
internal components.
NO flow must be continually evaluated during Nitric Oxide administration to ensure accurate dosing.
NO cannot be administered during the calibration process.
Do not sterilize or disinfect with power connected.
Allow unit to dry thoroughly before use. Immediate use after exposure to excessive cleaning agents
such as Isopropyl Alcohol can affect sensor performance.
Do not push on LCD display.
The VESA 75 mount is required for mounting during transport.
Should there be a sudden need to change therapy tanks; a second tank should always be purged and
ready for immediate use. Perform the purge procedure immediately upon installation of a new
regulator.
The flush procedure must be performed each time NO therapy is started. This includes initial therapy
starts, tank changes, and re-starting therapy after NO has been off.
Содержание AeroNO 2.0
Страница 1: ......
Страница 3: ...This page intentionally left blank ...
Страница 6: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 3 SCLAIM MER ...
Страница 7: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 4 SCLAIM MER ...
Страница 8: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 5 SCLAIM MER ...
Страница 9: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 6 SCLAIM MER ...
Страница 10: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 7 SCLAIM MER ...
Страница 11: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 8 SCLAIM MER ...
Страница 12: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 9 SCLAIM MER ...
Страница 26: ...Par rt No 715 008 2 13 2 14 88 Rev B Physical D VESA Mou Dimensions unting Patte and Cente ern 23 er of Mass ...
Страница 78: ...Par rt No 715 008 88 Rev B 5 12 3 Re 5 12 4 Fr ear Interna ront Intern al Layout al Layout 75 ...
Страница 94: ......
Страница 95: ...Par rt No 715 008 7 3 88 Rev B Nitric Oxid de Sensor S Specificatio 92 ns ...
Страница 96: ...Par rt No 715 008 88 Rev B 93 ...
Страница 97: ...Par rt No 715 008 7 4 88 Rev B NO2 Senso or Specifica ations 94 ...
Страница 98: ...Par rt No 715 008 88 Rev B 95 ...
Страница 99: ...Par rt No 715 008 7 5 88 Rev B Oxygen Se ensor Spec cifications 96 ...
Страница 105: ......