Part No. 715-0088, Rev. B
- 72 -
5.11.7.6. NO
2
Safety Shutoff Alarm
Check the safety shutoff alarm function by ensuring that:
5.11.7.6.1. When the displayed NO
2
value exceeds the value set on
the HI alarm limit by 1 ppm (or more), the audible alarm
activates and the AeroNOx 2.0 display alternates
between the message NITRIC OFF and the measured
NO and NO
2
values. Set HI alarm setting by 1 ppm or
more above sampled NO
2
value and observe.
Record
this by placing a checkmark in the corresponding box
for Safety Shutoff: Alarm on the AeroNOx 2.0 Gas
Testing Record Form.
Ensure that there is no gas flow
from the AeroNOx 2.0 delivery line. This can be verified
using an AeroNOx 2.0 delivery line and a cup of water to
check for bubbles.
Record this by placing a checkmark
in the corresponding box for Safety Shutoff: Test on
the AeroNOx 2.0 Gas Testing Record Form.
5.11.7.6.2. While in safety shutoff alarm condition, simultaneously
pressing and holding the NO and NO
2
alarm silence
buttons for 5 seconds enables the safety shutoff override.
The display should read SAFETY OFF once the safety
shutoff has been overridden. Ensure that the AeroNOx
2.0 display now alternates between the message
SAFETY OFF and the measured NO and NO
2
values.
Since the safety system has been deactivated, ensure that
there is gas flow from the AeroNOx 2.0 delivery line.
This can be verified using an AeroNOx 2.0 delivery line
and a cup of water to check for bubbles.
Record this by
placing a checkmark in the corresponding box for
Safety Shutoff: Override Test on the AeroNOx 2.0
Gas Testing Record Form.
5.11.7.6.3. While the safety shutoff is overridden, power the AeroNOx
2.0 off then power the AeroNOx 2.0 back on. This re-
activates the safety shutoff and the AeroNOx 2.0
delivery and alarms should function as normal.
5.11.8.
Sign and date the AeroNOx 2.0 Gas Testing Record Form
and any other
accompanying documentation required.
The unit is now ready for QA Final Inspection, which should occur soon after to
prevent discrepancies in testing. A test battery should still be installed to
maintain bias achieved during initial calibration.
Содержание AeroNO 2.0
Страница 1: ......
Страница 3: ...This page intentionally left blank ...
Страница 6: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 3 SCLAIM MER ...
Страница 7: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 4 SCLAIM MER ...
Страница 8: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 5 SCLAIM MER ...
Страница 9: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 6 SCLAIM MER ...
Страница 10: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 7 SCLAIM MER ...
Страница 11: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 8 SCLAIM MER ...
Страница 12: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 9 SCLAIM MER ...
Страница 26: ...Par rt No 715 008 2 13 2 14 88 Rev B Physical D VESA Mou Dimensions unting Patte and Cente ern 23 er of Mass ...
Страница 78: ...Par rt No 715 008 88 Rev B 5 12 3 Re 5 12 4 Fr ear Interna ront Intern al Layout al Layout 75 ...
Страница 94: ......
Страница 95: ...Par rt No 715 008 7 3 88 Rev B Nitric Oxid de Sensor S Specificatio 92 ns ...
Страница 96: ...Par rt No 715 008 88 Rev B 93 ...
Страница 97: ...Par rt No 715 008 7 4 88 Rev B NO2 Senso or Specifica ations 94 ...
Страница 98: ...Par rt No 715 008 88 Rev B 95 ...
Страница 99: ...Par rt No 715 008 7 5 88 Rev B Oxygen Se ensor Spec cifications 96 ...
Страница 105: ......