Part No. 715-0088, Rev. B
- 59 -
5.9.2.13. Open the AeroNOx 2.0 flow control knob fully (CCW). If already fully
opened, do not force it further. Knob may turn 10+ times.
5.9.2.14. Connect Beswick (small quick disconnect) to AeroNOx 2.0 .
5.9.2.15. Turn on the AeroNOx 2.0 . Hold down the Alarm Silence and Back
buttons simultaneously for ~5 seconds to enter Safety Off mode.
This disables alarms and protections to allow for diagnostics to be
performed. If the battery is not installed, use the external 9 VDC supply.
5.9.2.16. Connect the delivery fitting and length of Tygon tubing to the output
fitting.
5.9.2.17. Verify flow is present.
5.9.2.18. Block the externally attached delivery tube. Note the delivery line
connection may pop off the fitting if the pressure is too high.
5.9.2.19. Close the shut-off valve.
5.9.2.20. Observe the low pressure gauge for 1 minute. If pressure decreases
visibly, check the AeroNOx 2.0 for a leak. Repair leak if detected and
repeat test until no leak observed.
5.9.2.21. The delivery side of the AeroNOx 2.0 is confirmed to be leak free.
Check box on test record.
5.9.2.22. Unblock the delivery tube. Verify that gas is flowing and that the
AeroNOx 2.0 displays NO (L/min) 2.00 L/min.
Check box on test
record.
5.9.2.23. Attach the delivery line to the external flowmeter. Note that the actual
flow should be much higher than 2.00 L/min due to the Porter valve
being fully opened, but the AeroNOx 2.0 only displays accurately up to
2.0 L/min. The external flowmeter should measure higher if it is capable
of a higher range.
5.9.2.24. Close the AeroNOx 2.0 flow control knob partially (CW) until external
flowmeter reads 2.00 L/min.
Record values
on test record table
.
5.9.2.25. Repeat measurements for nominal flow settings in table.
Record
values
.
5.9.2.26. Close the AeroNOx 2.0 flow control knob completely (CW) until
external flowmeter reads 0.01 L/min.
5.9.2.27. Submerge the free end of the Tygon tubing in a cup of water and make
sure there is no flow (bubbles). Verify the display NO (L/min) reads
0.01 L/min then
check box on test record.
5.9.2.28. Leak check completed and mass flow sensor check complete.
Check
box on test record,
Sign and Date.
5.9.2.29. Perform the Sample Rate Adjustment Procedure before closing the
case if already opened.
Содержание AeroNO 2.0
Страница 1: ......
Страница 3: ...This page intentionally left blank ...
Страница 6: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 3 SCLAIM MER ...
Страница 7: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 4 SCLAIM MER ...
Страница 8: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 5 SCLAIM MER ...
Страница 9: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 6 SCLAIM MER ...
Страница 10: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 7 SCLAIM MER ...
Страница 11: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 8 SCLAIM MER ...
Страница 12: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 9 SCLAIM MER ...
Страница 26: ...Par rt No 715 008 2 13 2 14 88 Rev B Physical D VESA Mou Dimensions unting Patte and Cente ern 23 er of Mass ...
Страница 78: ...Par rt No 715 008 88 Rev B 5 12 3 Re 5 12 4 Fr ear Interna ront Intern al Layout al Layout 75 ...
Страница 94: ......
Страница 95: ...Par rt No 715 008 7 3 88 Rev B Nitric Oxid de Sensor S Specificatio 92 ns ...
Страница 96: ...Par rt No 715 008 88 Rev B 93 ...
Страница 97: ...Par rt No 715 008 7 4 88 Rev B NO2 Senso or Specifica ations 94 ...
Страница 98: ...Par rt No 715 008 88 Rev B 95 ...
Страница 99: ...Par rt No 715 008 7 5 88 Rev B Oxygen Se ensor Spec cifications 96 ...
Страница 105: ......