Part No. 715-0088, Rev. B
- 61 -
5.10. AeroNOx 2.0 Sample Rate Adjustment Procedure
This document provides instructions for adjusting sample flow and the Sample Blocked
trip point for AeroNOx 2.0 devices. This procedure is to be performed for all production
units prior to their initial release, yearly for units returning from the field, and as part of
every servicing process. Changes in sample flow rate are difficult to detect with normal
device usage and a low sampling rate may cause slow or unstable sensor readings.
This procedure may be performed on an AeroNOx 2.0 with or without sensors. The
device must have a battery connected to function properly, even with external power
applied to the DC jack. If a unit under test passes, it does not have to be disassembled
and adjusted.
5.10.1.
Tools and Materials
ESD-protected workstation and grounded wrist strap
738-1853
AeroNOx 2.0 Sample/Delivery Kit
738-1930
Power Supply, Medical Grade, 40 W, 9 V, (731-9159 units)
738-1964
Power Supply, Medical Grade, 40 W, 9 V, Locking (731-0426
units)
300 mm - 0 - 300 mm water column manometer or equivalent pressure
meter capable of measuring 60 cm H2O or 5900 Pa
Gas flowmeter setup with precision metering valve
416-0010
Calibration Screwdriver, Single
4 mm hex wrench
Multi-meter with leads
5.10.2.
Instructions
NOTE:
All flow measurements are assumed to be at standard atmospheric
conditions (STP).
5.10.2.1.Connect power supply to the AeroNOx 2.0 under test, power the
AeroNOx 2.0 on, and allow the unit to run at least 1 minute before
attempting to set sample flow as described below.
5.10.2.2.Set up the flowmeter and precision metering valve as shown. Ensure
that both the flow meter and precision metering valves are fully open (fully
counterclockwise) to allow unrestricted flow, and adjust the zero on the
paper marker to the H
2
O level in the water column.
5.10.2.3.Connect the sample line to the AeroNOx 2.0 sample line inlet.
Содержание AeroNO 2.0
Страница 1: ......
Страница 3: ...This page intentionally left blank ...
Страница 6: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 3 SCLAIM MER ...
Страница 7: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 4 SCLAIM MER ...
Страница 8: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 5 SCLAIM MER ...
Страница 9: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 6 SCLAIM MER ...
Страница 10: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 7 SCLAIM MER ...
Страница 11: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 8 SCLAIM MER ...
Страница 12: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 9 SCLAIM MER ...
Страница 26: ...Par rt No 715 008 2 13 2 14 88 Rev B Physical D VESA Mou Dimensions unting Patte and Cente ern 23 er of Mass ...
Страница 78: ...Par rt No 715 008 88 Rev B 5 12 3 Re 5 12 4 Fr ear Interna ront Intern al Layout al Layout 75 ...
Страница 94: ......
Страница 95: ...Par rt No 715 008 7 3 88 Rev B Nitric Oxid de Sensor S Specificatio 92 ns ...
Страница 96: ...Par rt No 715 008 88 Rev B 93 ...
Страница 97: ...Par rt No 715 008 7 4 88 Rev B NO2 Senso or Specifica ations 94 ...
Страница 98: ...Par rt No 715 008 88 Rev B 95 ...
Страница 99: ...Par rt No 715 008 7 5 88 Rev B Oxygen Se ensor Spec cifications 96 ...
Страница 105: ......