Part No. 715-0088, Rev. B
- 86 -
6.2. Troubleshooting Other Symptoms and Solutions
The following section describes the most common symptoms that are likely to occur with
the AeroNOx 2.0 and suggests the possible solutions.
SYMPTOMS AND SOLUTIONS
1.
Instrument has been sitting for some time. It has not been connected to a charger and the
power switch does not turn the instrument on. The sample pump does not run.
Possible solutions:
1.
The battery is disconnected, or has become discharged.
Connect charger, cycle power, and see if proper operation
now occurs. Recharge battery for at least 7 hours.
2.
The unit has been subjected to a severe ESD occurrence.
Disconnect the AC power supply and remove and disconnect
battery for a few seconds. Reconnect battery and check
operation.
3.
Replace battery if it fails to charge.
2.
Instrument has been sitting for some time. It has not been connected to a charger. When
charger was connected and the unit turned on, very high NO values are indicated on the LCD
and the alarms are sounding.
Possible solutions:
1.
The battery has become deeply discharged and the
instrument has been unable to maintain the bias voltage on
the NO sensor. Connect charger and allow battery to charge
and the instrument to bias the NO sensor for at least 48 hours
before attempting to calibrate the instrument and take
measurements.
2.
If battery fails to take a charge, replace the battery.
3.
Oxygen analyzer will not calibrate to room air of 21% O
2
.
Possible solution:
1.
Check the output of the O
2
sensor for approximately 10 mV
output in room air. If < 9 mV, replace sensor.
2.
Check cable and connectors connecting the sensor to the
circuit board.
4.
Oxygen analyzer will not calibrate to 100% O
2
.
Possible solutions:
1.
Check the output of the O
2
sensor for approximately 10 mV
output in room air. If < 9 mV, replace sensor. 13 mV is
expected at 100% O
2
.
2.
If sensor checks OK above, check plumbing, tubing, and
connections for leaks in the system. Ambient air is probably
being drawn into the system. Repair leaks.
3.
Check cable and connectors connecting the sensor to the
circuit board.
Содержание AeroNO 2.0
Страница 1: ......
Страница 3: ...This page intentionally left blank ...
Страница 6: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 3 SCLAIM MER ...
Страница 7: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 4 SCLAIM MER ...
Страница 8: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 5 SCLAIM MER ...
Страница 9: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 6 SCLAIM MER ...
Страница 10: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 7 SCLAIM MER ...
Страница 11: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 8 SCLAIM MER ...
Страница 12: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 9 SCLAIM MER ...
Страница 26: ...Par rt No 715 008 2 13 2 14 88 Rev B Physical D VESA Mou Dimensions unting Patte and Cente ern 23 er of Mass ...
Страница 78: ...Par rt No 715 008 88 Rev B 5 12 3 Re 5 12 4 Fr ear Interna ront Intern al Layout al Layout 75 ...
Страница 94: ......
Страница 95: ...Par rt No 715 008 7 3 88 Rev B Nitric Oxid de Sensor S Specificatio 92 ns ...
Страница 96: ...Par rt No 715 008 88 Rev B 93 ...
Страница 97: ...Par rt No 715 008 7 4 88 Rev B NO2 Senso or Specifica ations 94 ...
Страница 98: ...Par rt No 715 008 88 Rev B 95 ...
Страница 99: ...Par rt No 715 008 7 5 88 Rev B Oxygen Se ensor Spec cifications 96 ...
Страница 105: ......