Part No. 715-0088, Rev. B
- 67 -
5.11.3.
Battery / Charge Indicator Check
5.11.3.1. Ensure that the battery symbol is full when external power is
temporarily unplugged. Record this by placing a checkmark in the
corresponding box on the AeroNOx 2.0 Gas Testing Record Form.
5.11.3.2. Ensure that the charge symbol is displayed, green LED lights up on
front and on back when connected to the 9 V power supply specified
for use with the AeroNOx 2.0 .
Record this by placing a
checkmark in the corresponding box on the AeroNOx 2.0 Gas
Testing Record Form.
5.11.4.
Zero Calibration
5.11.4.1. Allow the AeroNOx 2.0 to sample room air for at least 2 minutes.
Cycle power, then hold down the Back button to enter calibration.
Press zero cal, then press it again to hold to zero. O
2
should now
read 21%. NO and NO
2
are zeroed.
Record this by placing
checkmarks in the corresponding boxes on the AeroNOx 2.0
Gas Testing Record Form.
5.11.5.
Oxygen Analyzer Calibration
5.11.5.1. Connect the 731-0127 (or equivalent) Oxygen hose from the
Compressed Oxygen Cylinder with regulator to the 731-9130 (or
equivalent) Oxygen flowmeter input.
5.11.5.2. Open the Compressed Oxygen Cylinder valve and adjust the pressure
regulator to 50 5 psi and flow through the Oxygen flowmeter to ~ 1
L/min.
5.11.5.3. Connect the O
2
dedicated Calibration Circuit (738-1850) from the
Oxygen flowmeter output to the AeroNOx 2.0 sample port.
5.11.5.4. Enter calibration mode again and select High Cal, then O
2
. Allow
the AeroNOx 2.0 to sample 100% oxygen for 2 minutes. Use arrows
to set O
2
value to 100%.
Record this by placing a checkmark in
the corresponding box on the AeroNOx 2.0 Gas Testing Record
Form.
If the analyzer is still out of specification or will not hold a
100% measurement, the analyzer is nonconforming. Follow standard
operating procedures for nonconforming product.
5.11.5.5. Disconnect the O
2
Calibration Circuit and close the Oxygen Cylinder
valve.
NOTE:
The use of a low volume shutoff valve is permitted, but
each gas must be on an isolated flow path to prevent
gas cross contamination.
Содержание AeroNO 2.0
Страница 1: ......
Страница 3: ...This page intentionally left blank ...
Страница 6: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 3 SCLAIM MER ...
Страница 7: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 4 SCLAIM MER ...
Страница 8: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 5 SCLAIM MER ...
Страница 9: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 6 SCLAIM MER ...
Страница 10: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 7 SCLAIM MER ...
Страница 11: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 8 SCLAIM MER ...
Страница 12: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 9 SCLAIM MER ...
Страница 26: ...Par rt No 715 008 2 13 2 14 88 Rev B Physical D VESA Mou Dimensions unting Patte and Cente ern 23 er of Mass ...
Страница 78: ...Par rt No 715 008 88 Rev B 5 12 3 Re 5 12 4 Fr ear Interna ront Intern al Layout al Layout 75 ...
Страница 94: ......
Страница 95: ...Par rt No 715 008 7 3 88 Rev B Nitric Oxid de Sensor S Specificatio 92 ns ...
Страница 96: ...Par rt No 715 008 88 Rev B 93 ...
Страница 97: ...Par rt No 715 008 7 4 88 Rev B NO2 Senso or Specifica ations 94 ...
Страница 98: ...Par rt No 715 008 88 Rev B 95 ...
Страница 99: ...Par rt No 715 008 7 5 88 Rev B Oxygen Se ensor Spec cifications 96 ...
Страница 105: ......