Part No. 715-0088, Rev. B
- 68 -
5.11.6.
Nitric Oxide (NO) Gas Testing
5.11.6.1. NO Calibration
5.11.6.1.1. Ensure that the AeroNOx 2.0 complete with sensors to
be calibrated has been plugged in for at least 48 hours
prior to being gas tested. This will ensure that the NO
sensor has biased properly.
NOTE:
This time can be reduced if the NO sensor
has been biased correctly for at least 48
hours.
5.11.6.1.2. Power on the AeroNOx 2.0 and enter calibration mode if
not already there.
5.11.6.1.3. Note that the AeroNOx 2.0 accuracy is 1 ppm for NO,
so displayed values need to be within this specification.
5.11.6.1.4. Purge the EPA Protocol Gas, 80 ppm NO, Balance
Nitrogen Cylinder.
5.11.6.1.5. Connect the 80 ppm NO dedicated 201-0055 Female
CGA 660 to Male CGA 625 Adapter to the 80 ppm NO
Cylinder and the 80 ppm NO dedicated 731-9141
Calibration Regulator. Connect the 80 ppm NO dedicated
738-1850 Calibration Circuit to the 80 ppm NO Calibration
Regulator. Open the 80 ppm NO Cylinder valve for 30
seconds to flush the circuit.
5.11.6.1.6. Connect the sample line plug of the 80 ppm NO
Calibration Circuit to the AeroNO2.0 sample port.
5.11.6.1.7. Select High Cal, then NO. Allow the AeroNOx 2.0 to
sample NO for approximately 5 minutes or until stable.
Use arrows to set NO value to match bottle calibration
(e.g. 80 ppm).
Record bottle cal value and high cal set
value on the AeroNOx 2.0 Gas Testing Record Form.
If the set value maxes out without matching the bottle
calibration or will not hold a stable NO measurement, the
analyzer is nonconforming. Follow standard operating
procedures for nonconforming product.
5.11.6.1.8. Disconnect the 80 ppm NO Calibration circuit and close
the 80 ppm NO Cylinder valve. Open the Oxygen
Cylinder valve and connect the O
2
Calibration Circuit to
the AeroNOx 2.0 sample port. Ensure that the displayed
NO value returns to 0-1 ppm and O
2
goes to 98-100%.
Record Re-check Zero
value on Gas Testing Record
Form for NO.
Содержание AeroNO 2.0
Страница 1: ......
Страница 3: ...This page intentionally left blank ...
Страница 6: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 3 SCLAIM MER ...
Страница 7: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 4 SCLAIM MER ...
Страница 8: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 5 SCLAIM MER ...
Страница 9: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 6 SCLAIM MER ...
Страница 10: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 7 SCLAIM MER ...
Страница 11: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 8 SCLAIM MER ...
Страница 12: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 9 SCLAIM MER ...
Страница 26: ...Par rt No 715 008 2 13 2 14 88 Rev B Physical D VESA Mou Dimensions unting Patte and Cente ern 23 er of Mass ...
Страница 78: ...Par rt No 715 008 88 Rev B 5 12 3 Re 5 12 4 Fr ear Interna ront Intern al Layout al Layout 75 ...
Страница 94: ......
Страница 95: ...Par rt No 715 008 7 3 88 Rev B Nitric Oxid de Sensor S Specificatio 92 ns ...
Страница 96: ...Par rt No 715 008 88 Rev B 93 ...
Страница 97: ...Par rt No 715 008 7 4 88 Rev B NO2 Senso or Specifica ations 94 ...
Страница 98: ...Par rt No 715 008 88 Rev B 95 ...
Страница 99: ...Par rt No 715 008 7 5 88 Rev B Oxygen Se ensor Spec cifications 96 ...
Страница 105: ......