Part No. 715-0088, Rev. B
- 57 -
5.9. AeroNOx 2.0 MFM Flow Check/Leak Test Procedure
This document contains detailed instructions for checking the MFM (mass flow meter)
flow and for ensuring that there are no leaks in the delivery circuit plumbing or around the
stainless steel fittings in an AeroNOx 2.0 . Nitrogen is used as a test flow gas. This
procedure is to be performed as part of the AeroNOx 2.0 production process prior to
initial release of the product, at least once annually for units returning from the field, and
as part of every servicing process because:
MFM zero may drift with device age.
Leaks may be introduced inadvertently by sharp jarring of the device.
Leaks may be introduced inadvertently during servicing of the gas delivery circuit.
Leaks may develop due to embrittlement of tubing or stressed connnections.
This procedure may be performed on an AeroNOx 2.0 with or without sensors. The
device must have a battery connected to function properly, even with external power
applied to the DC jack.
5.9.1.
Tools and Materials
ESD-protected workstation and grounded wrist strap
4 mm hex wrench
738-1930
Power Supply, Medical Grade, 40 W, 9 V (for 731-9159 units)
738-1964
Power Supply, Medical Grade, 40 W, 9 V, Locking (for 731-
0426 units)
888-0115
Battery Assembly, SLA, 6 V, 4.5 Ah
888-0013
Battery Assembly, LiFePO
4
, 6.4 V, 4.5 Ah
Potentiometer adjusting tool (e.g. P/N 416-0010 - Calibration Screwdriver,
Single)
738-1845
Coupling Body, 1/8 Barb
374-1630
Tubing, Tygon, 1/8 B. 1/16 W. 1/4 OD approximately 3
feet
Braided stainless steel hose with Swagelok QC4 male quick connect fitting
and Beswick miniature quick disconnect (e.g. P/N 738-1862 - AeroNOx 2.0
NO Supply Hose, 6). Use of an adapter to an existing benchtop setup is
acceptable.
Compressed nitrogen delivery gas cylinder with regulator, high side
pressure gauge, and low side pressure gauge capable of measuring 30-70
psi, shut-off valve, and female Swagelok quick connect fitting (P/N 278-
1002 - Quick Connect, Female, 1/8 Bulkhead).
Flow meter with range 0 to 2.0 L/min (e.g. Omega FMA1816A-ST)
Содержание AeroNO 2.0
Страница 1: ......
Страница 3: ...This page intentionally left blank ...
Страница 6: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 3 SCLAIM MER ...
Страница 7: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 4 SCLAIM MER ...
Страница 8: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 5 SCLAIM MER ...
Страница 9: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 6 SCLAIM MER ...
Страница 10: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 7 SCLAIM MER ...
Страница 11: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 8 SCLAIM MER ...
Страница 12: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 9 SCLAIM MER ...
Страница 26: ...Par rt No 715 008 2 13 2 14 88 Rev B Physical D VESA Mou Dimensions unting Patte and Cente ern 23 er of Mass ...
Страница 78: ...Par rt No 715 008 88 Rev B 5 12 3 Re 5 12 4 Fr ear Interna ront Intern al Layout al Layout 75 ...
Страница 94: ......
Страница 95: ...Par rt No 715 008 7 3 88 Rev B Nitric Oxid de Sensor S Specificatio 92 ns ...
Страница 96: ...Par rt No 715 008 88 Rev B 93 ...
Страница 97: ...Par rt No 715 008 7 4 88 Rev B NO2 Senso or Specifica ations 94 ...
Страница 98: ...Par rt No 715 008 88 Rev B 95 ...
Страница 99: ...Par rt No 715 008 7 5 88 Rev B Oxygen Se ensor Spec cifications 96 ...
Страница 105: ......