Part No. 715-0088, Rev. B
- 69 -
5.11.6.1.9. Disconnect the O
2
Calibration circuit and close the
Oxygen Cylinder valve. Open the 80 ppm NO Cylinder
valve and connect the 80 ppm NO Calibration Circuit to
the AeroNOx 2.0 sample port. Wait for at least 2
minutes for the gas concentrations to stabilize. Ensure
the displayed NO value returns to the established high cal
value.
Record
this value on the corresponding line for
Re-check High Span
on the AeroNOx 2.0 Gas
Testing Record Form. If it does not, repeat high cal once
or twice. If these steps have already been repeated, it is
probable that the NO sensor may have expired. Remove
the AeroNOx 2.0 from testing and follow approved
procedures for non-conforming product.
5.11.6.2. NO Mid-Range Reference Testing
5.11.6.2.1. Disconnect the 80 ppm NO Calibration Circuit and close
the 80 ppm NO Cylinder valve.
5.11.6.2.2. Purge the EPA Protocol Gas, 45 ppm NO, Balance
Nitrogen Cylinder.
5.11.6.2.3. Connect the 45 ppm NO dedicated 201-0055 Female
CGA 660 to Male CGA 625 Adapter to the 45 ppm NO
Cylinder and the 45 ppm NO dedicated 731-9141
Calibration Regulator. Connect the 45 ppm NO dedicated
738-1850 Calibration Circuit to the 45 ppm NO Calibration
Regulator. Open the 45 ppm NO Cylinder valve for 30
seconds to flush the circuit.
5.11.6.2.4. Connect the sample line plug of the 45 ppm NO
Calibration Circuit to the AeroNOx 2.0 sample port.
Allow the AeroNOx 2.0 to sample NO for approximately
5 minutes or until stable.
Record bottle cal value and
AeroNOx2.0 NO reading on the AeroNOx 2.0 Gas
Testing Record Form.
If the AeroNOx 2.0 NO reading
does not match the bottle cal value within 1 ppm or will
not hold a stable NO measurement, the analyzer is
nonconforming. Follow standard operating procedures for
nonconforming product.
5.11.7.
Nitrogen Dioxide (NO
2
) Gas Testing
5.11.7.1. NO
2
Calibration
Repeat the steps for NO Calibration (Section 5.11.6.1.) substituting
NO
2
in place of NO noting the following differences:
5.11.7.1.1. Note that the AeroNOx 2.0 accuracy is 0.1 ppm for
NO
2
, so displayed values need to be within this
specification.
5.11.7.1.2. Use EPA Protocol Gas, 9 ppm Nitrogen Dioxide, Balance
Air for the High Cal gas.
5.11.7.1.3. Use the 9 ppm NO
2
dedicated CGA Adapter, Calibration
Circuit, and Calibration Regulator.
Содержание AeroNO 2.0
Страница 1: ......
Страница 3: ...This page intentionally left blank ...
Страница 6: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 3 SCLAIM MER ...
Страница 7: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 4 SCLAIM MER ...
Страница 8: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 5 SCLAIM MER ...
Страница 9: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 6 SCLAIM MER ...
Страница 10: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 7 SCLAIM MER ...
Страница 11: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 8 SCLAIM MER ...
Страница 12: ...Par rt No 715 008 88 Rev B LA ANGUAG GE DIS 9 SCLAIM MER ...
Страница 26: ...Par rt No 715 008 2 13 2 14 88 Rev B Physical D VESA Mou Dimensions unting Patte and Cente ern 23 er of Mass ...
Страница 78: ...Par rt No 715 008 88 Rev B 5 12 3 Re 5 12 4 Fr ear Interna ront Intern al Layout al Layout 75 ...
Страница 94: ......
Страница 95: ...Par rt No 715 008 7 3 88 Rev B Nitric Oxid de Sensor S Specificatio 92 ns ...
Страница 96: ...Par rt No 715 008 88 Rev B 93 ...
Страница 97: ...Par rt No 715 008 7 4 88 Rev B NO2 Senso or Specifica ations 94 ...
Страница 98: ...Par rt No 715 008 88 Rev B 95 ...
Страница 99: ...Par rt No 715 008 7 5 88 Rev B Oxygen Se ensor Spec cifications 96 ...
Страница 105: ......