Chapter 4
4–2
MiniMate Plus Operator Manual
The monitor powers up the sensors and displays the average and maximum difference read from
the sensors. This allows you to test the sensor by stimulating it (i.e., shaking a transducer or
blowing into a microphone) and observing the response.
4.3. Setups Key
The Setups key accesses the following items. The BlastWare III software is the easiest method to
setup the monitor. The following setups can be changed in the field using the monitor alone.
4.3.1. Sample Rate
The MiniMate Plus offers a number of sample rates in the advanced mode including 512, 1024,
2048, 4096, 8192, 16384 samples per second (32768 and 65536 optional). Histogram record
mode supports the 1024, 2048 and 4096 sample rates only (the 4096 sample rate can only be used
with a maximum of four enabled channels). The following table summarizes the sample rates and
the channels that may be used. Note that some sample rates operate on specific channels only.
Sample Rate Availability
Sample Rate
Channel 1 Only Channels 1 – 2 Channels 1 – 4 Channels 1 – 8
65536
available
– – –
32768 available
available –
–
16384
available available available
–
8192
available available available available
4096
available available available available
2048
available available available available
1024
available available available available
512
available available available available
“–” indicates not available
To choose a sample rate, press the
Setups
key with the Main Window displayed. Use the
Enter
key to scroll through the windows until the Sample Rate Window appears. Press the
Right
Arrow
key to scroll through the available sample rates. Press the
Enter
key to select a sample rate when
it appears on the display.
Note:
Eight channel monitors have been optimized for sample rates up to 65536 samples per
second. Four channel monitors may exhibit noise at sample rates greater than 16384
samples per second. Please contact the manufacturer for further information.
4.3.2. Channel Range
The Channel Range specifies the measurement limits of each channel. Two ranges are available:
Sensitive (Gain = 8) and Normal (Gain = 1). Each channel may be set to a different range.
To choose a channel range, press the
Setups
key with the Main Window displayed. Use the
Enter
key to scroll through the windows until the Channel Range Window appears. Use the
Up
or
Right
Arrow
keys to display the required range. Press the
Enter
key to save the range and display the
next channel. Continue until you have set the channel range for all the recording channels.
Содержание 716A0401
Страница 1: ......
Страница 6: ...iv MiniMate Plus Operator Manual Notes...
Страница 12: ...x MiniMate Plus Operator Manual Notes...
Страница 16: ...Chapter 1 1 4 MiniMate Plus Operator Manual 1 4 MiniMate Plus Keys...
Страница 66: ...Chapter 5 5 16 MiniMate Plus Operator Manual Notes...
Страница 71: ...Appendix MiniMate Plus Operator Manual A 1 Appendix...
Страница 72: ...Appendix A 2 MiniMate Plus Operator Manual...
Страница 81: ...Appendix MiniMate Plus Operator Manual A 11 Notes...
Страница 82: ...Appendix A 12 MiniMate Plus Operator Manual...