Compliance Module
MiniMate Plus Operator Manual
3–23
3.12. Test Key
Pressing the
Test
key executes Sensorcheck and tests the MiniMate Plus and its sensors.
Sensorcheck performs a two stage test. In the first stage, the program displays the MiniMate Plus
serial number, software version, the total amount of memory installed, the amount of memory
available to store events, and the number of events presently stored in memory.
R 4.0 V 4.0
Operating System
Release and Version
Figure 3.17 The Sensorcheck Window.
The second stage tests each geophone within the standard transducer and the microphone. The
program also tests the operation of the MiniMate Plus itself and the transducer and microphone
connecting cables. The message “All Channels Working. Press Test to Review Setups”
indicates the MiniMate Plus passed the Sensorcheck.
3.13. Option Key
The
Option
key, when used in combination with other keys, provides the following functions.
3.13.1. Option Key + Setups Key
Pressing the
Option
key and the
Setups
key at the same time accesses advanced functions
including reset factory defaults, auto window, auxiliary trigger, backlight on time, power saving
timeout, and monitoring LCD cycle.
a. Reset Factory Defaults
This function returns all the MiniMate Plus setups to the original factory defaults. If you choose to
reset the monitor, all of your setups will lost and replaced with the factory defaults. You can
reprogram your setups again anytime you want.
WARNING:
Resetting the Factory Defaults deletes your existing setups. Your setups will be
lost.
To reset factory defaults, press the
Option
key and the
Setups
key together when the Main
Window is displayed. The Reset Factory Defaults Window appears. Press the
Enter
key to
continue or the
Cancel
key to exit. Press and hold the
Enter
key to reset the factory defaults or
press the
Cancel
key to move to the next window.
b. Auto Window
When the trigger level is set to Auto, the Auto Window determines the length of time the
MiniMate Plus continues to record for after an event’s activity falls below the trigger level. This
ensures that an events activity, particularly air blast, is recorded. For example, if you plan to
record an event at a long distance from the source, set the Auto Window to a large value to
compensate for the time required for the air blast to travel from the source to the monitor’s
microphone. If you set the value to a small value, the monitor may stop recording after the seismic
activity reaches it but before the air blast reaches it. The Auto Window values range from one to
nine seconds.
To set the Auto Window, press the
Option
key and the
Setups
key together when the Main
Window is displayed. Press the
Cancel
key to exit the Reset Factory Defaults Window. Press the
Содержание 716A0401
Страница 1: ......
Страница 6: ...iv MiniMate Plus Operator Manual Notes...
Страница 12: ...x MiniMate Plus Operator Manual Notes...
Страница 16: ...Chapter 1 1 4 MiniMate Plus Operator Manual 1 4 MiniMate Plus Keys...
Страница 66: ...Chapter 5 5 16 MiniMate Plus Operator Manual Notes...
Страница 71: ...Appendix MiniMate Plus Operator Manual A 1 Appendix...
Страница 72: ...Appendix A 2 MiniMate Plus Operator Manual...
Страница 81: ...Appendix MiniMate Plus Operator Manual A 11 Notes...
Страница 82: ...Appendix A 12 MiniMate Plus Operator Manual...