Chapter 3
3–18
MiniMate Plus Operator Manual
To set the Geo Alarm Level, press the
Right Arrow
key and the
Setups
key together. Press the
Enter
key repeatedly to scroll through the windows until the Geo Alarm Level Window appears.
Use the
Up
and
Right Arrow
keys to set the alarm level. Press the
Enter
key to save your setting.
3.9.10. Mic Alarm Level
(Compliance Module Only)
The Mic Alarm Level controls the triggering of an optional remote alarm (711A1401). On eight
channel monitors, the Mic Alarm Level operates on channels one to four only.
The remote alarm sounds two alarms: warning and alarm. The warning alarm sounds when event
activity exceeds the microphone trigger level. The Mic Alarm Level can be set to a higher level
than the microphone trigger level. This allows for the triggering of a second alarm at a higher
level. See the Geo Alarm Level section, above, for an example.
To set the Mic Alarm Level, press the
Right Arrow
key and the
Setups
key together. Press the
Enter
key repeatedly to scroll through the windows until the Mic Alarm Level Window appears.
Use the
Up
and
Right Arrow
keys to set the alarm level. Press the
Enter
key to save your setting.
3.9.11. Serial Connection
This sets the MiniMate Plus to communicate with a computer using the BlastMate III PC
Connecting Cable (712A2301) or via a modem over telephone lines or radio frequencies. You
must set the MiniMate Plus to the proper serial connection before starting communications. To
communicate with a computer using the connecting cable, choose the Direct setting. To
communicate using modems, choose the Via Modem setting. Set the baud rate on the monitor to
match the baud rate of your computer or modem.
To choose the serial connection, press the
Right Arrow
key and the
Setups
key together. Press
the
Enter
key to scroll the windows until the Serial Connection Window appears. Press
Right
Arrow
key repeatedly to display the two choices. Select Direct if you are connecting to a
computer with the BlastMate cable or Via Modem if you are using modems. Press the
Enter
key
to select a choice when it appears on the display.
3.9.12. Direct Baud Rate or Modem Baud Rate
This sets the baud rate for the serial connection method chosen in the previous step, either direct
or via modem. For direct, choose a baud rate of 38400. For modem communications, choose a
baud rate supported by the modems you are using.
3.9.13. Call Home
This setting is used to setup the monitor for use with the Auto Call Home feature of the BlastWare
III software. With Auto Call Home, your monitor can be setup to automatically download results
either after an event, or at specified times. You must have a modem and phone line connected to
the unit in order to use this feature. In addition, you must setup Auto Call Home within BlastWare
III on your PC. Please refer to the BlastWare III Operator Manual (p/n 714U0301) for information
on how to configure Auto Call Home and setup modems for use with a monitor.
To turn Call Home on, select Enabled. If you do not wish to use this feature, select Disabled.
To set Call Home, press the
Right Arrow
key and the
Setups
key together when the Main
Window is displayed. Press the
Enter
key to scroll through the windows until the Call Home:
Window appears. Press the
Up
or
Right Arrow
key to display either Enabled or Disabled. Press
the
Enter
key to save your setting.
Note
: Auto Call Home is supported by MiniMate Plus units with a serial number of 6000 or
higher.
Содержание 716A0401
Страница 1: ......
Страница 6: ...iv MiniMate Plus Operator Manual Notes...
Страница 12: ...x MiniMate Plus Operator Manual Notes...
Страница 16: ...Chapter 1 1 4 MiniMate Plus Operator Manual 1 4 MiniMate Plus Keys...
Страница 66: ...Chapter 5 5 16 MiniMate Plus Operator Manual Notes...
Страница 71: ...Appendix MiniMate Plus Operator Manual A 1 Appendix...
Страница 72: ...Appendix A 2 MiniMate Plus Operator Manual...
Страница 81: ...Appendix MiniMate Plus Operator Manual A 11 Notes...
Страница 82: ...Appendix A 12 MiniMate Plus Operator Manual...