8 | Cleaning and maintenance
INFICON
84 / 100
Pernicka-700H-CHLD-Operating-instructions-jin76aen1-04-(1911)
Screw (2x)
Guide groove
Earth tongue
Electronic device
Analyzer
QME 220
QMA 220
Fig. 55: Mount the QME 220 electronic device onto the QMA 220 analyzer
Connect electrical connections to the IO 220 module
► Reestablish the electrical connections with the IO 220 module: Power supply,
Ethernet cable, analog I/O-cable, digital I/O-cable.
8.4 Maintenance of the internal calibration leak
8.4.1 Perform calibration
Carry out the calibration of the device according to the maintenance plan, see
87]". Perform calibration, see "Calibrating [
58]".
Required spare parts
None
Required tools
None
8.4.2 Recalibration of the calibration leak by VTI
Carry out the maintenance work according to the maintenance plan, see "Maintenance
Required spare parts
None
Required tools
Please contact the INFICON service
1
Remove the calibration leak. Please contact the INFICON service.