7 | Warning and error messages
INFICON
76 / 100
Pernicka-700H-CHLD-Operating-instructions-jin76aen1-04-(1911)
Message
Possible causes
Solution
Turning on the emission
and SEM FAILED and
was aborted!
SEM amplifier dirty or
defective.
Please contact the INFICON
service.
FAILURE: Emission and
SEM are OFF!
HV pressure is too
high.
Wait until the pressure is small
enough to turn on the SEM
amplifier.
FAILURE: Air pressure
for valve switching is too
low!
Air pressure for the
valves is less than 7
bar.
Check that the air pressure is
greater than 7 bar.
Open the pressure valves in the
supply lines.
Gradually increase the pressure
each by 0.5 bar.
Problems
Situation
Possible causes
Solution
The device cannot be
switched on.
The power cord for the
power supply is not
plugged into the
device.
Plug the supplied power cord into
the POWER LINE INPUT plug
and connect the power cord to
the power supply.
Red X in the system
diagram ("Measure"
window) at the point "Air
pressure".
The pressure of the
compressed air supply
is less than 6.9 bar.
Check that the air pressure is
greater than 7 bar.
Open the pressure valves in the
supply lines.
Gradually increase the pressure
each by 0.5 bar.
The lid of the test
chamber cannot be
opened.
The test chamber is
under vacuum.
In the "Pernicka 700H"
measurement program, press the
"Measure" tab and then the
"Load/Unload" button.
Ensure that the pressure of the
purge gas supply is between 69
mbar and 71 mbar.
Ensure that the pressure of the
compressed air supply is between
6.9 bar and 7.59 bar.
Hissing or stuttering
sound from the back of
the device.
The pressure of the
compressed air supply
is greater than the
factory-set limit for the
pressure relief valve.
Check that the pressure of the
compressed air supply is less
than 9 (consistency!) bar. Lower
the pressure by 0.2 bar.