INFICON
Warning and error messages | 7
Pernicka-700H-CHLD-Operating-instructions-jin76aen1-04-(1911)
75 / 100
Message
Possible causes
Solution
Change to the other cathode in
the "Pernicka 700H"
measurement program. See
Emission current > +/-
20 %
HV pressure is too
high.
Cathode is aged.
Change to the other cathode in
the "Pernicka 700H"
measurement program. See
RF temperature high >
65 °C ... 80 °C
Filter mat clogged in
the CHLD-Sensor.
Fan in the CHLD-
Sensor defective.
RF generator in the
CHLD-Sensor is
incorrectly calibrated.
Change the filter mat in the
CHLD-Sensor. See "Check or
81]".
Change the fan in the CHLD-
Sensor. Please contact the
INFICON service.
Error messages
Message
Possible causes
Solution
External protection
triggered
HV pressure is too
high.
Check the tightness of the valves.
Emission Error
HV pressure is too
high.
Check the tightness of the valves.
Filament 1 defect
Cathode 1 is defective. Change to the cathode 2 in the
"Pernicka 700H" measurement
program. See "Select cathode in
Please contact the INFICON
service. The second cathode is
equivalent to the first and typically
allows for a period of about a
month of device operation.
Turning on Analyzer
FAILED!
HV pressure is too
high.
Both cathodes are
defective.
Internal connection
problem in the CHLD-
Sensor.
Replace both cathodes in the
CHLD-Sensor. Please contact the
INFICON service.
An Error occurred (long
text)
Internal Windows
communication error.
Start the measurement program
"Pernicka 700H" once a week.