8 | Cleaning and maintenance
INFICON
80 / 100
Pernicka-700H-CHLD-Operating-instructions-jin76aen1-04-(1911)
The maintenance interval according to the maintenance schedule (see
1
Perform an acoustic check on the fan function on the QME 220 electronic unit.
2
If you do not hear any fan noise, change the fan. Please contact the INFICON
service.
3
If you hear fan noise, check the filter mat. See "Check or change the filter mat
81]".
The temperature in the QME 220 electronic unit is too high
The temperature in the QME 220 electronic unit, is normally in the range of 20 °C ...
25 °C above the ambient temperature.
1
To open the "Tools” window, press the tab "Tools".
2
Tools > Tune Up.
Display
QME Temp"
"
Fig. 50: CHLD-Sensor: Check temperature of the QME 220 electronic unit
3
The current temperature of the QME 220 electronics device of the CHLD-Sensor
is displayed in the "QME Temp" field. Check.
4
If the temperature in the QME 220 electronics device is 35 °C above the
ambient temperature, perform an acoustic check on the fan function on the QME
220 electronics device.
5
If you do not hear any fan noise, change the fan. Please contact the INFICON
service.
6
If you hear fan noise, check the filter mat. See "Check or change the filter mat
81]".