5 | Installation
INFICON
30 / 100
Pernicka-700H-CHLD-Operating-instructions-jin76aen1-04-(1911)
WARNING
Danger from liquids, moisture and electricity
Moisture entering the device can lead to personal injury due to electric shocks as well
as damage to property due to short circuiting.
► Only operate the device in dry environments and only in buildings.
► Operate the device away from sources of liquid and moisture.
► Place the device where you can always reach the mains plug.
► Do not operate the device standing water and do not let even a drop of water or
other liquids on the device.
► Prevent the device from coming into contact with bases, acids and solvents.
CAUTION
There is a danger uncontrolled rolling away
When on an incline the device can roll away uncontrollably.
► Fix the device by latching the front rollers.
NOTICE
Material damage from overheated device
The device and above all the cryogenic pump compressor become warm during
operation and can overheat without sufficient ventilation.
► Please note the technical specifications.
► Ensure sufficient ventilation, especially on the ventilation slots on the left and right
of the device: Free space in the front, rear and sides of at least 10 cm for the
device; 1 m for the cryogenic pump compressor.
► Keep heat sources away from the device and from the cryogenic pump compressor.
Avoiding measurement errors due to helium sources in the device
environment
The helium concentration in the region of the Pernicka 700H CHLD may differ only
insignificantly from the natural helium concentration of 5 ppm. The detection limit of
the Pernicka 700H CHLD is strongly impaired by higher helium concentrations.