Pernicka-700H-CHLD-Operating-instructions-jin76aen1-04-(1911)
61 / 100
1
To turn the device on, set the power switch to ON (see "Switching the device on
ð
The device has a power-on delay and is switched on after approx. 20 s.
ð
The computer and monitor are automatically turned on and started.
ð
The operating system is started and the "Measure" window is displayed by
default.
Idle time
► You must wait for about 2 hours after switching on the device. During this time, the
cryogenic pump is cooled down to a temperature < 20 K. If System pumped and
Ready" appears in the "Mini" or "Measure" in the yellow highlighted message
window, you can proceed to the next step.
2
Perform a calibration. See "Calibrating [
58]".
3
If you want to use only helium as tracer gas, close the flow control valve's
knurled screw (see "Select single-gas [
57]"). If you want to use several tracer
gases, open the flow control valve's knurled screw (see "Select multi-gas
4
Use a suitable tool such as a plastic tweezer to place the test object in the test
chamber and close the lid of the test chamber.
5
Make sure that the correct basic settings and the settings for the current
measurement are carried out. See "Basic settings [
42]".
6
Make sure that the desired method is set see "Select method [
7
Set the setpoint for the gross leak measurement, see "Set the setpoint for gross
8
Make sure that calibration takes place daily. See "Calibrating [
9
Ensure that the regeneration of the cryogenic pump is carried out once a day.
10
To start the measurement process manually, see "Prepare for manual start
6.5.2 Prepare for manual start
You can start the measurement process manually with the round connector inserted.