Equipment Connections
The Ecotec E3000 is equipped with three electrical control connectors. The electrical
connections (head phone, I/O port and RS232) are located on the rear of the Ecotec
E3000 directly next to the socket for the mains cable.
7.1
I/O Port (Control Inputs and Outputs)
Through this connection some functions of the Ecotec E3000 can be controlled
externally or measurement data or the status of the Ecotec E3000 may be
communicated to external equipment.
Through relay change over contacts the trigger levels as well as the operating mode
(Ready) of the Ecotec E3000 may be monitored.
Warning
For all contacts of the I/O Port a maximum voltage of 60 V DC or 25 V AC must not
be exceeded or reached to ground or ground equipment conductors.
According to the type of in- or outputs lower voltages had to be accepted. For this,
please refer to the information given in the responding chapters.
Fig. 7-1 Default Pin Assignment
Содержание Ecotec E3000
Страница 2: ...0 2 kina22eIVZ fm Technical Handbook 1206 ...
Страница 13: ...General Information 1 3 kina22e chapter 1 fm technical handbook 1206 ...
Страница 30: ...1 20 General Information kina22e chapter 1 fm technical handbook 1206 ...
Страница 38: ...2 8 How the Instrument Works kina22e chapter 2 fm technical handbook 1206 ...
Страница 58: ...3 20 Operation of the Ecotec E3000 kina22e chapter 3 fm technical handbook 1206 ...
Страница 98: ...4 40 Equipment Settings kina22e chapter 4 fm technical handbook 1206 ...
Страница 106: ...6 8 Ecotec E3000 Messages kina22e chapter 6 fm technical handbook 1206 ...
Страница 133: ...Maintenance 8 18 kina22e chapter 8 fm technical handbook 1206 ...
Страница 143: ...Annex 9 10 kina22e chapter 9 fm technical handbook 1206 9 2 Declaration of Conformity ...